Multi-Purpose Volumetric Pump
The AFF represents the next generation of medical infusion technology, offering the "all in one" solution
by unifying various dedicated-treatment infusion pumps and multi-set capability.
This revolutionary device is based on advanced technology that integrates various modes of infusion
operations such as PCA, TPN, adult/infant flow rates and a wide range of other features found only on
most expensive high-end pumps. Together these features comprise a comprehensive, high-value/low-
cost solution to health care providers, and offer an outstanding features per dollar relationship to
the end-user.
The advanced technology of the AFF allows for its instant recalibration to suit a wide variety of standard
IV sets, previously a costly and time-consuming process.
The AFF is designed to function in primary applications that include intensive care units, critical care
units, neonatal units, operating procedures, chemotherapy, anesthesiology and pain management and
is suitable for home-care use.
Definitive’s revolutionary infusion technology:
- Comprehensive integration of various
dedicated-treatment infusion pumps
- Superior accuracy and safety features
- Multi-set capability
- Highly cost effective and versatile
- Dedicated adult/infant flow rates
- Large LCD display
- Ergonomic graphic user interface for incredibly easy use
- IV line guide mechanis
- Intuitive default parameters setting
- Sturdy construction offers lifelong dependability
- Multi-language option
- Internal software upgradeable
- Alarm history log
- Unique clamp mechanism preventing flow when set is removed from the pump (optional)
- Medication/Weight calaulation (optional)
- Remote monitoring (optional)
AFF Technical Specifications
| Infusion |
| Flow Rate |
1-1,200 ml/ hr in 1ml increments (Adult) 0.1-120 ml/ hr in 0.1ml increments (Infant) |
| Accuracy* |
±5% using recommended sets |
| Volume Range |
1-9,999 ml in 1 ml increments (Adult) 0.1-999.9 ml in 0.1 ml increments (Infant) |
| KVO |
1 or 5 ml/hr (Adult), 0.1 ml/hr (Infant) |
| Fast Bolus |
1-1,200 ml/ hr, Volume: 1-999 ml (Adult) 1-120 ml/ hr, Volume: 1-99.9 ml (Infant) |
| Mechanical |
| Pumping Mechanism |
Peristaltic |
| Ultrasonic Air Detector |
Accumulative: 25-500μl/ 15 min, Adjustable threshold |
| Drop Sensor (optional) |
External |
| Upstream Pressure Sensor |
Adjustable sensitivity |
| Downstream Pressure Sensor |
Adjustable sensitivity |
| Pressure Sensitivity Range |
0.2-1.2 bar (downstream) |
| Door Clamp |
IV set free flow protection when door opens |
| IV Set Requirements |
Suitable for use with Infusion Pump: OD = 3.5-5mm, 55-72 Shore A |
| Dimensions |
120 x 195 x 240 mm |
| Weight (net) |
3.0 Kg (approx.) |
| Construction |
Plastic- double insulated |
| Alarms |
Pressure (occlusion: upstream/ downstream), near end, infusion complete, KVO, high flow, low flow, drop sensor, battery low, no power, air in line, door open, system error, alarm time counter, error-sensors, real time battery, clamp missing (optional) |
| Special Features |
| PCA |
Patient Control Analgesia |
| TPN |
Total Parenteral Nutrition |
| Volume Over Time |
Automatically calculates the flow rate according to the infusion volume and duration |
| Second Bag |
Enables a dual bag infusion with different parameters |
| KVO (Keep Vein Open) |
Continues to infuse at a minimal rate to keep the vein open |
| ABS (Anti Bolus System) |
Releases pressure due to bolus buildup |
| Fast Bolus |
Operator initiated accelerated infusion |
| Purge |
Clears IV line of air bubbles |
| PC Connection |
RS232 socket enables maintenance and parameters settings using an external PC software interface |
| Setting Protection |
Secure settings password protected (for authorized staff only) |
| Alarm Log |
Records crucial alarms |
| Languages |
Programmable user language interface |
| RT Clock |
Displays real time clock |
| Tube Guide Mechanism |
Keeping IV set in peristaltic movement |
| Anti Free Flow Clamp Mechanism (optional) |
A unique innovative clamping mechanism, additional to door clamp, that prevents flow when the IV set is removed
from the pump |
| Medication/ Weight (optional) |
Programming infusion parameters using medication and patient data |
| Remote Alarm Indication (optional) |
Remote indication for alarms when manual intervention is required |
| Power |
| Power Supply Requirements |
100-240 VAC, 50/60 Hz, 2 Amp. max. |
| Battery |
Rechargeable Ni-Metal Hydride 7.2V 3.7 Ahr |
| Battery Operation (at 125 ml/hr) |
~10 hrs |
| Memory Backup Battery |
Lithium CR 1620 |
|