Mennen Medical's Enmove 1200 is a sophisticated, portable
bedside patient monitor, providing a highly efficient solution
for the varying needs of today's medical institutions.
This monitor is versatile, easily adaptable to suit any
medical application and patient profile and enables
the remote viewing of other Enmove monitors as
well as central station connectivity. This innovative
device provides long term tabular and numerical
trend and is available with a variety of configurations:
from basic non invasive to multi-parameter
monitoring, covering all medical requirements,
including gas analysis.
Ease of use:
Intuitive operation and functionality,
with adaptable display, user-friendly interface design,
large sized fonts, adjustable Vital Signs sequence
and colors to fit the user’s requirements and work
routines.
Flexibility:
Easily adaptable to a wide variety of patient
environments with specified requirements. Full patient
customization. Wireless connectivity to Local Area Network.
Increased Clinical Value: Maximal simultaneous parameter presentation, including 12 leads,
numeric and graphic ECG display, waveforms and numerical vital signs for the whole spectrum
of patient monitoring displayed on the screen, long-term follow-up of vital signs trend in both
numeric and graphic format as well as medical calculations.
Cost Efficiency:
Low maintenance and spare parts costs.
Clinical Features:
Parameters
? 5 - 12 lead ECG
? Respiration
? NIBP
? SpO2 (Nelcor/Masimo)
● Temperature
● EtCO2 Microstream
● Mainstream Etco2
● Analog Output for ECG and IBP
● 2 BP
● Cardiac Output
● Gas Analysis (Artema)
Event Strips: Event strips of all waveforms and alarms at the time of the event.
Graphic Trend: 96 hours Trend panel presentation of graphic trends, including up to eight vital
signs with up to one minute resolution.
Numerical Trend: 96 hours Numerical values of vital signs displayed in up to one minute resolution,
including alarms and event marking.
Arrhythmia Analysis: Continuous ECG waveform analysis, based on QRS algorithm including
12 leads analysis.
ST Analysis: Simultaneous and continuous ST analysis of displayed ECG leads with ST alarm.
All-inclusive ECG lead display: Simultaneous display of all ECG leads.
Drug Titration and Calculation: Drug concentrate, infusion rate and injection amount calculations
performed, according to defined clinical parameters.
Wireless Network: Fixed compact flash for memory card or wireless LAN card.
Dimension and Weight
● Dimension: 318mm (W) x 264mm (H) x 152mm (D)
● Weight: <7.5 kg
Operation Environment
● Power: AC100-240V (10%), 50/60 Hz (3Hz), 140VA
● Temperature: 0-400C
● Humidity: 15-95% non-condensing
Patient Range
Neonate, pediatric and adult patients
Performance Specifications
● Display: 12.1” color TFT
Rolling and refreshing waveform display
● Resolution: 800x600
Multi displays selectable, including:
- Standard display
- Large-font Display
- Freeze display
- Alarm limit display
- Multi lead and ECG simultaneous display
- Bed-to-bed view display
● Trace: 8 waveforms
● Sweep speed: 6.25mm/s, 12.5mm/s,
25mm/s, 50mm/s
● Indicator: Alarm indicator
- Power indicator
- QRS beep and alarm sound
- Working indicator light
- QRS beep and alarm sound
● Interface: Parameter cable interface
- AC Power input socket
- Network interface
- External, VGA interface for connection of and
alternate display
● Li-ion Battery: Rechargeable
(For 2 pieces): Maximum 6.5 hours for charging,
5 hours continuous working
● Lead Acid Battery: Rechargeable
(For 2 pieces): Maximum 12 hours for charging,
2 hours continuous working
● Trend time: 1-96 hours
● Alarm: User adjustable High and Low limits, 3
level audible and visual alarm
● Recorder: Built-in thermal array, 2 channels
- Record mode: manual, on alarm, time defined
- Paper width: 50mm
- Record speed: 25mm/s, 50mm/s
ECG
5-lead and 3-lead selectable, 12 lead
(including 3/5 lead) optional
● Input:
- 10 lead wire cable: RA; LA; RL; LL; V1-V6 or
R; L; N; F; C1-C6
- 5-lead: RA; LA; RL; LL; V or R; L; N; F; C
- 3-lead: RA; LA; LL or R; L; F
● Lead selection:
- 12 lead: I; II; III; avR; avL; avF; V1-V6
- 5-lead: I; II; III; avR; avL; avF; V (n)
- 3-lead: I; II; III
● ECG waveform: 2 channels
● Gain selection: x0.125; x0.25; x0.5; x1;
x2(mm/mV); auto
● Sweep speed: 12.5mm/s, 25mm/s, 50mm/s
● Heart Rate range: Adult: 15-300bpm;
Pediatric/Neonatal: 15-350bpm
● Accuracy: ±1bpm or ±1%, whichever is greater
● Resolution: 1bpm
● Filter:
- Diagnostic mode: 0.05-100Hz or 0.05-150Hz
(optional 12 lead)
- Monitoring mode: 0.5-40Hz
- Surgical mode:1-20Hz
● Protection: Withstand 4000VAC/50Hz voltage
in isolation. Against electrosurgical interference
and defibrillation
● Scaling signal: 1mv+/- 5%
● Alarm range: 15-350bpm
● S-T detection:
- Measurement range: -2.0mV~2.0mV
- Alarm range: -2.0mV~2.0mV
● Arrhythmia analysis: YES
● Alarm: YES, audible and visual alarm, alarm
events recallable
Respiration
● Method: Thoracic impedance
● Operation modes: Auto/Manual
● Measurement range: Adult: 0-120rpm,
Pediatric/Neonatal: 0-150rpm
● Apnea alarm: YES
● Alarm: YES, audible and visual alarm, alarm
events recallable
NIBP
● Method: Automatic Oscillometric
● Operation modes:
Manual/Automatic/Continuous
● Auto measure time: Adjustable
● Measurement unit: mmHg/kPa selectable
● Measurement types: Systolic, Diastolic, Mean
● Measurement range:
Range of systolic pressure:
- Adult: 40-270mmHg
- Pediatric: 40-200mmHg
- Neonatal: 40-135mmHg
Range of diastolic pressure:
- Adult: 10-210mmHg
- Pediatric: 10-150mmHg
- Neonatal: 10-95mmHg
Range of mean pressure:
- Adult: 20-230mmHg
- Pediatric: 20-165mmHg
- Neonatal: 20-110mmHg
● Accuracy: The Mean error shall be less than +/-5
mmHg and standard deviation shall be less than 8 mmHg
● Over-pressure protection: double safety
protection
● Alarm: Systolic, Diastolic, and Mean
Temperature
Measurement range: 0-50°C
Resolution: 0.1°C
Accuracy: ±0.1°C (not including probe)
Channel: Dual-channel
Alarm range: 0-50°C
SpO2(Standard)
● Measurement range: 0-100%
● Resolution: 1%
● Accuracy:
+/- 2 %( 70-100% Adult/Pediatric, non-motion);
+/- 3 %( 70-100% neonate, non-motion);
+/- 3%( 70-100% Adult/Pediatric/Neonate, motion);
0-69% unspecified
● Alarm range: 0-100%
● Pulse rate:
- Range: 20-254bpm
- Resolution: 1bpm
- Accuracy: ±3bpm (non-motion), ±5bpm (motion)
● Alarm range: 20-254bpm
Masimo SpO2
● Measurement range: 1-100%
● Resolution: 1%
● Accuracy:
+/- 2 %( 70-100% Adult/Pediatric, non-motion);
+/- 3 %( 70-100% Neonate, non-motion);
+/- 3%( 70-100% Adult/Pediatric/Neonate, motion);
0-69% unspecified
● Alarm range: 0-100%
● Pulse rate:
- Range: 25-254bpm
- Resolution: 1bpm
- Accuracy: ±3bpm (non-motion), ±5bpm (motion)
● Alarm range: 25-254bpm
NELLCORE SpO2
● Measurement range: 1-100%
● Resolution: 1%
● Accuracy:
+/- 2 %( 70-100%, MAX-A, MAX-AL, MAX-N, MAXP,
MAX-I AND MAX-FAST sensors);
+/- 2.5 %( 70-100% OxiCliq A, OxiCliq N, OxiCliq
P and OxiCliq I sensors);
+/- 3%( 70-100% D-YS, DS-100A, OXI-A/N AND
OXI P/I sensors);
+/- 3.5%( 70-100% MAX-R, D-YSE AND D-YSPD
sensors);
0-69% unspecified
● Alarm range: 0-100%
● Pulse rate:
- Range: 20-300bpm
- Resolution: 1bpm
- Accuracy: ±3bpm (20-250bpm),
(251-300bpm) unspecified
● Alarm range: 25-250bpm
IBP
● Measurement range: -50~300mmHg
● Channel: 2 channels
● Pressure transducer: Sensitivity: 5 mV/V/mmHg
● Impedance range: >300
● Transducer sites: ART, PA, CVP, RAP, LAP, ICP , CPP
● Resolution: 1mmHg
● Accuracy: ±1mmHg or ±2%, whichever is
greater (exclusive of transducers) |